Effect of stimulation parameters on entrainment

Open AccessPublished:September 21, 2022DOI:https://doi.org/10.1016/j.hroo.2022.09.012

      Keywords

      Key Findings
      • Increasing stimulation intensity may shorten the postpacing interval (PPI).
      • Pacing rate may affect PPI.
      • Number of stimuli does not affect PPI.
      The postpacing interval (PPI) may be longer (>30 ms) or even shorter (<0 ms) than the tachycardia cycle length (TCL) when entrainment pacing is performed within the isthmus of the circuit.
      • Kaiser D.W.
      • Hsia H.H.
      • Dubin A.M.
      • et al.
      The precise timing of tachycardia entrainment is determined by the postpacing interval, the tachycardia cycle length, and the pacing rate: theoretical insights and practical applications.
      ,
      • Johner N.
      • Shah D.C.
      • Jousset F.
      • et al.
      Electrophysiological and anatomical correlates of sites with postpacing intervals shorter than tachycardia cycle length in atypical atrial flutter.
      In this study, we aimed to evaluate whether stimulation parameters can affect the PPI in typical (cavotricuspid isthmus–dependent) atrial flutter (AFL).
      Twenty-one patients with typical AFL were included in the study. Each patient provided written informed consent to the study protocol, which was approved by the Institutional Ethics Committee (Research No. 2020FYYX032). The research reported in this paper adhered to Helsinki Declaration.
      After confirmation of a typical AFL, entrainment pacing with different parameters was performed. Pacing trains were performed with different numbers of pacing stimuli (16, 24, and 32 basic S1S1 stimulations), different stimulation intensities (5, 8, and 10 mA), and different pacing rates (10 ms less than TCL, 20 ms less than TCL, 30 ms less than TCL).
      Continuous variables (PPI–TCL) were compared using the paired t test or Wilcoxon signed rank test. P <.05 was considered significant. All analyses were performed using SPSS Version 17.0 (SPSS Inc., Chicago, IL).
      There was no significant difference in PPI–CL between 16 and 24 basic S1S1 stimulations (12.0 ± 11.1 ms vs 12.2 ± 11.0 ms, respectively; P = .55) and between 24 and 32 basic S1S1 stimulations (12.2 ± 11.0 ms vs 11.4 ± 10.4 ms, respectively; P = .05). This finding indicates that the number of pacing stimuli does not affect PPI. There was a significant difference in PPI–CL between stimulation intensities of 5 and 8 mA (11.4 ± 10.5 ms vs 7.1 ± 8.7 ms; P = .01) and between stimulation intensities of 8 and 10 mA (7.1 ± 8.7 ms vs 3.20 ± 7.6 ms; P = .01). An example is shown in Figures 1A to 1C. Figures 1D and 1E show the possible mechanisms. When stimulation intensity increases, more of the far-field tissue may be captured because of the virtual electrode and the downstream wavefront forming farther from the pacing site and then returning to the pacing site earlier. As a result, PPI is shortened. Despite the existence of variability in PPI,
      • Derejko P.
      • Podziemski P.
      • Zebrowski J.J.
      • et al.
      Effect of the restitution properties of cardiac tissue on the repeatability of entrainment mapping response.
      we can clearly see this tendency. Although the difference observed when the entrainment parameters are changed can consist of only a few milliseconds, it could lead to misclassification of an entrained site when the PPI–TCL interval is close to the threshold of 30 ms. The possible mechanism and an example is shown in Figures 1F to 1I.
      Figure thumbnail gr1
      Figure 1A: Entrainment pacing in the isthmus at a stimulus intensity of 5 mA with a postpacing interval (PPI) of 255 ms. The tachycardia cycle length (TCL) was 255 ms, and pacing rate was 245 ms. B: Entrainment pacing in the isthmus at a stimulus intensity of 8 mA with a PPI of 246 ms. C: Entrainment pacing in the isthmus at a stimulus intensity of 10 mA with a PPI of 235 ms. D: Pacing at lower intensity. The virtual electrode (oval) is next to the pacing site. E: Pacing at higher intensity. The virtual electrode is farther away. F: If pacing near the scar or the area of conduction block (eg, eustachian ridge [ER]), the pacing virtual electrode may capture one or the other side of the barrier side. The circuit may be changed and the PPI changed significantly. G: Pacing at lower intensity. PPI was 253 ms, TCL was 220 ms, and pacing rate was 210 ms. H: Pacing at higher intensity. PPI was 213 ms. Note the double potentials recorded at the pacing electrode. I: Three-dimensional mapping showing typical atrial flutter. Yellow tags represent the His bundle; red tags represent ablation sites; black tags represent scars; and blue tags represent double potentials. CS = coronary sinus; CT = crista terminalis; IVC = inferior vena cava; SVC = superior vena cava.
      There was a significant difference in PPI-CL between pacing with 10 ms less than TCL and 20 ms less than TCL (9.3 ± 5.9 ms vs 10.6 ±6.2 ms; P <.001), and between pacing with 20 ms less than TCL and 30 ms less than TCL (10.6 ± 6.2 ms vs 9.2 ±7.3 ms, respectively; P = .04). Faster pacing may lead to decremental conduction or concealed conduction, which may affect PPI.
      • Vollmann D.
      • Stevenson W.G.
      • Lüthje L.
      • et al.
      Misleading long post-pacing interval after entrainment of typical atrial flutter from the cavotricuspid isthmus.
      In conclusion, PPI is related to stimulation intensity and rate. Increasing stimulation intensity may shorten PPI, and the fast pacing rate also may affect PPI.

      Funding Sources

      This work was supported by Chongqing Medical Scientific Research Project (Joint Project of Chongqing Health Commission and Science and Technology Bureau) (2020FYYX032).

      Disclosures

      The authors have no conflicts of interest to disclose.

      Authorship

      All authors attest they meet the current ICMJE criteria for authorship.

      Patient Consent

      Each patient provided written informed consent to the study protocol.Ethics Statement and Acknowledging: The study protocol was approved by the Institutional Ethics Committee (Research No. 2020FYYX032). The research reported in this paper adhered to the Helsinki Declaration.

      References

        • Kaiser D.W.
        • Hsia H.H.
        • Dubin A.M.
        • et al.
        The precise timing of tachycardia entrainment is determined by the postpacing interval, the tachycardia cycle length, and the pacing rate: theoretical insights and practical applications.
        Heart Rhythm. 2016; 13: 695-703
        • Johner N.
        • Shah D.C.
        • Jousset F.
        • et al.
        Electrophysiological and anatomical correlates of sites with postpacing intervals shorter than tachycardia cycle length in atypical atrial flutter.
        Circ Arrhythm Electrophysiol. 2019; 12e006955
        • Derejko P.
        • Podziemski P.
        • Zebrowski J.J.
        • et al.
        Effect of the restitution properties of cardiac tissue on the repeatability of entrainment mapping response.
        Circ Arrhythm Electrophysiol. 2014; 7: 497-504
        • Vollmann D.
        • Stevenson W.G.
        • Lüthje L.
        • et al.
        Misleading long post-pacing interval after entrainment of typical atrial flutter from the cavotricuspid isthmus.
        J Am Coll Cardiol. 2012; 59: 819-824